FDA Adverse Event
Malfunction
Summary report: N
RIATA ST ACTIVE FIXATION
MDR report key: 2962766
·
Received February 14, 2013
Report
- Report Number
- 2938836-2013-00107
- Event Type
- Malfunction
- Date Received
- February 14, 2013
- Date of Event
- December 7, 2012
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. (B)(4). NO COMPLAINT RECEIVED WITH RETURN OF DEVICE. FAILURE (EVENT) OBSERVED DURING ANALYSIS. AN ENTIRE LEAD WAS RETURNED CUT IN TWO PIECES AT 14.7 CM FROM THE CONNECTOR PIN. EXTERNAL INSULATION ABRASION WAS FOUND AT 43.5-43.7 CM FROM THE DISTAL TIP CONSISTENT WITH LEAD FRICTION TO THE ICD CAN. THE ETFE COATING WAS INTACT AT THE SAME LOCATION.
Description of Event or Problem · 1
THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 65700 | RIATA ST ACTIVE FIXATION | DEFIBRILLATION LEAD | NVY | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 7000/65 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |