FDA Adverse Event
Malfunction
Summary report: N
HOMECHOICE PRO
MDR report key: 2962628
·
Received February 14, 2013
Report
- Report Number
- 1416980-2013-03877
- Event Type
- Malfunction
- Date Received
- February 14, 2013
- Date of Event
- January 14, 2013
- Report Date
- February 4, 2013
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE DEVICE WAS RETURNED TO BAXTER, AND THE EVALUATION IS COMPLETE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT. THE REPORTED DIFFICULTY OF AN IIPV EVENT WAS CONFIRMED THROUGH AN EVENT HISTORY LOG REVIEW. THE CAUSE WAS UNDETERMINED. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP WILL BE SENT.
Description of Event or Problem · 1
A HIGH DRAIN ERROR 101 (NIGHT DRAIN #1) ALARM WAS IDENTIFIED IN THE LOG OF A RETURNED HOMECHOICE DEVICE. THE ALARM OCCURRED ON (B)(6) 2013, 08:43:15. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 64705 | HOMECHOICE PRO | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - LARGO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |