FDA Adverse Event Malfunction Summary report: N

ILAB ULTRASOUND IMAGING SYSTEM

MDR report key: 2962543 · Received February 14, 2013

Report

Report Number
2134265-2013-00645
Event Type
Malfunction
Date Received
February 14, 2013
Report Date
January 18, 2013
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
OBJ
PMA / PMN Number
K051679
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MANUFACTURER: THE PRODUCT WAS RECEIVED IN GOOD CONDITION WITH NO VISIBLE EXTERNAL DAMAGE OR DEFECTS OBSERVED. THE MDU-5 WAS INSTALLED INTO A GOLD STANDARD SYSTEM AND TESTED FOR FUNCTIONALITY. THE UNIT UNDERWENT A 4 HOUR BURN-IN WITHOUT ANY FAILURES OBSERVED. THE UNIT MEETS SPECIFICATIONS. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE WAS NOT CONFIRMED. (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATED BY MFR: THE DEVICE HAS NOT BEEN RECEIVED FOR ANALYSIS. UPON RECEIPT AND COMPLETION OF THE FAILURE ANALYSIS OF THE COMPLAINT DEVICE, IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MEDWATCH WILL BE FILED. (B)(4).

Description of Event or Problem · 1

SAME CASE AS MDR ID # 2134265-2013-0064, 2134265-2013-00647. IT WAS REPORTED THAT DURING PERCUTANEOUS CORONARY INTERVENTION (PCI) PROCEDURE, THE AUTOMATIC PULLBACK OF THE ILAB MOTORDRIVE STOPPED. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT STATUS IS STABLE.

Description of Event or Problem · 1

SAME CASE AS MDR ID # 2134265-2013-0064, 2134265-2013-00647. IT WAS REPORTED THAT DURING PERCUTANEOUS CORONARY INTERVENTION (PCI) PROCEDURE, THE AUTOMATIC PULLBACK OF THE ILAB MOTORDRIVE STOPPED. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT STATUS IS STABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
65930 ILAB ULTRASOUND IMAGING SYSTEM CATHETER, ULTRASOUND, INTRAVASCULAR OBJ BOSTON SCIENTIFIC - MAPLE GROVE H749ILAB100CART0

Patients

Seq Age Sex Outcome Treatment
1