ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-01802
- Event Type
- Malfunction
- Date Received
- February 14, 2013
- Report Date
- January 20, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME.
DEVICE EVALUATION FOLLOW-UP #1 (B)(4): THE DEVICE WAS RETURNED TO ANIMAS AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING RESULTS: UNUSUAL POWERS ON RESETS ARE OBSERVED IN THE BLACK BOX DATA. BATTERY CAP THREADS ARE INTACT. NO CRACK WAS OBSERVED TO THE BATTERY COMPARTMENT. THE BATTERY CAP IS ABLE TO SECURELY FIT TO THE PUMP. CAP CONTACTS MEASUREMENTS WERE TAKEN AND FOUND:WITHIN SPECIFICATION. BATTERY CAP WAS TIGHTEN UNTIL FLUSH AGAINST PUMP AND SLOWLY UNSCREWED ½ TURN WITH NO REBOOTS. THE BATTERY CAP IS ABLE TO MAINTAIN AN ELECTRICAL CONNECTION. THE PUMP POWERS UP WITH AUDIBLE SOUND AND VIBRATION, BUT THE DISPLAY IS BLANK: UNABLE TO VERIFY ALL STEPS TO INVESTIGATE INTERMITTENT POWER COMPLAINT. THE COVER WAS POPPED OFF; MOISTURE CORROSION WAS FOUND ON THE DISPLAY SCREEN FLEX/CONNECTOR AND ON THE PCB. UNRELATED TO THIS COMPLAINT, THE AUDIO BOLUS BUTTON COVER WAS OBSERVED TO BE DETACHED AND MISSING, UNABLE TO TAKE THE AUDIBLE TEST READING.
ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (INTERMITTENT POWER) ISSUE. THE PUMP REPORTEDLY POWERED OFF WITHOUT USER INTERVENTION. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 64910 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 10 YR |