FDA Adverse Event
Malfunction
Summary report: N
CORE IMPACTION DRILL
MDR report key: 2962131
·
Received January 24, 2013
Report
- Report Number
- 1811755-2013-90129
- Event Type
- Malfunction
- Date Received
- January 24, 2013
- Date of Event
- December 26, 2012
- Report Date
- December 27, 2012
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- DZJ
- PMA / PMN Number
- K040369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
ADD'L INFO WILL BE SUBMITTED ONCE THE QUALITY INVESTIGATION HAS BEEN PERFORMED. THE REASON FOR THE SERIALIZATION STARTING WITH 90XXX IS TO PROVIDE CLARIFICATION FOLLOWING A CHANGE IN STRYKER'S ELECTRONIC COMPLAINT SYSTEMS.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE AT THE USER FACILITY, THE DRILL WAS OVERHEATING. BACK-UP EQUIPMENT WAS USED TO COMPLETE THE PROCEDURE. NO DELAY, NO ADVERSE CONSEQUENCES AND NO MEDICAL INTERVENTION WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 34953 | CORE IMPACTION DRILL | DZJ | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |