FDA Adverse Event Malfunction Summary report: N

AU480 CHEMISTRY SYSTEM

MDR report key: 2961967 · Received February 14, 2013

Report

Report Number
9612296-2013-00011
Event Type
Malfunction
Date Received
February 14, 2013
Date of Event
January 18, 2013
Report Date
January 22, 2013
Manufacturer
BECKMAN COULTER
Product Code
JGS
PMA / PMN Number
K981743
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

NO SERVICE WAS DISPATCHED FOR THIS EVENT. NO FURTHER NA ISSUES HAVE BEEN REPORTED AT THIS TIME. CUSTOMER AGREED THAT THE NA CALIBRATION SLOPE SHIFT NOTED ON (B)(6) 2013, WAS THE CAUSE OF THE LOWERED NA RESULTS. FAILURE MODE OF THIS EVENT WAS SHIFTED CALIBRATION SLOPE RECOVERY.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER (BEC) TO REPORT A LOW SODIUM (NA) RECOVERY ON MULTIPLE PATIENT SAMPLES TESTED ON THE AU480 CHEMISTRY SYSTEM ON (B)(6) 2013. THE CUSTOMER STATED THAT NA WAS CALIBRATED FOR THE DAY AND QC WAS PERFORMED FOR THAT 24 HOUR PERIOD. QC WAS NOTED TO BE WITHIN LABORATORY ACCEPTABLE RANGES, BUT ON THE LOW SIDE. CUSTOMER THEN PROCEEDED TO RUN PATIENT SAMPLES. ON MONDAY, (B)(6) 2013, THE NA DAILY CALIBRATION WAS PERFORMED, AND THE CUSTOMER NOTICED THAT THE CALIBRATION SLOPE AND QC RECOVERY RETURNED TO THE LEVELS PREVIOUSLY NOTED BEFORE THE DROP ON (B)(6) 2013. THE CUSTOMER THEN RERAN ALL THE PATIENT SAMPLES FROM THE (B)(6) 2013 RUN AND DISCOVERED THAT MOST NA RESULTS WERE NOW ABOUT 10 UNITS HIGHER. TWENTY FIVE (25) AMENDED REPORTS WERE ISSUED. PATIENT RESULTS ARE PROVIDED IN FILE ATTACHMENT SECTION OF THIS REPORT. PER THE CUSTOMER, THERE HAVE BEEN NO REPORTS OF ANY EFFECT TO PATIENTS IN CONNECTION TO THE LOW NA RESULTS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
65456 AU480 CHEMISTRY SYSTEM AU480 CHEMISTRY ANALYZER JGS BECKMAN COULTER AU481-02E WITH ISE NA

Patients

Seq Age Sex Outcome Treatment
1