FDA Adverse Event Malfunction Summary report: N

INFUSOR

MDR report key: 2961799 · Received February 14, 2013

Report

Report Number
1416980-2013-03770
Event Type
Malfunction
Date Received
February 14, 2013
Date of Event
January 24, 2013
Report Date
January 25, 2013
Manufacturer
BAXTER HEALTHCARE - IRVINE
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: BAXTER RECEIVED ONE SAMPLE CONTAINING NO FLUID IN THE RESERVOIR. A PATIENT CONTROL MODULE (PCM) WAS CONNECTED. VISUAL EXAMINATION OF THE SAMPLE SHOWED NO SIGN OF PHYSICAL ABNORMALITY. AN ACCURACY FLOW TEST WAS PERFORMED ON THE SAMPLE FOR 19.2 HOURS. THE INFUSOR PRODUCED FUNCTIONAL RESULTS WITHIN THE SPECIFICATION RANGE; THE DEVICE PERFORMED AS EXPECTED. THE REPORTED CONDITION OF AN OVERINFUSION WAS NOT CONFIRMED. NO ROOT CAUSE WAS IDENTIFIED, AS NO EVIDENCE OF PRODUCT MALFUNCTION WAS OBSERVED. NO REPAIR WAS DONE, AS THIS IS A SINGLE-USE DEVICE WHICH WILL BE DISCARDED. NO OTHER OBSERVATIONS WERE NOTED ON THE UNIT. ADDITIONAL INFORMATION: THE BATCH REVIEW REVEALED THAT ALL OF THE ACCEPTANCE CRITERIA WERE MET TO RELEASE THE LOT. THERE WERE NO NON-CONFORMANCES, FAILURES, REWORK, OR DEVIATIONS RELATED TO THE LOT. THERE WERE NO CHANGES TO SPECIFICATIONS, TEST METHODS, PROCESS, EQUIPMENT, OR RAW MATERIALS THAT COULD BE ASSOCIATED WITH THE REPORTED CONDITION. THIS PRODUCT IS NOT DISTRIBUTED IN THE US AND DOES NOT HAVE A 510(K) NUMBER.

Description of Event or Problem · 1

BAXTER (B)(4) RECEIVED A REPORT OF AN BASAL/BOLUS INFUSOR THAT OVERINFUSED DURING PATIENT THERAPY. THE DEVICE WAS FILLED WITH SALINE, FENTANYL CITRATE AND DROPERIDOL. THE ACTUAL FLOW RATE WAS 40ML/9.75HR. THERE WAS PATIENT INVOLVEMENT, BUT THERE WAS NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
64643 INFUSOR PUMP, INFUSION FRN BAXTER HEALTHCARE - IRVINE 12F038

Patients

Seq Age Sex Outcome Treatment
1 DROPERIDOL| FENTANYL CITRATE| SALINE