FDA Adverse Event Malfunction Summary report: N

MEDFUSION 3500 SYRINGE INFUSION PUMP

MDR report key: 2961753 · Received January 16, 2013

Report

Report Number
2183502-2012-00673
Event Type
Malfunction
Date Received
January 16, 2013
Date of Event
December 17, 2012
Report Date
October 15, 2013
Manufacturer
SMITHS MEDICAL MD
Product Code
FRN
PMA / PMN Number
K040899
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NH, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE SUSPECT DEVICE WAS RETURNED AND EVALUATED. TESTING INDICATED THAT THE POSITION POTENTIOMETER WAS NON-FUNCTIONAL DUE TO A BROKEN TAB. THE POSITION POTENTIOMETER WAS REPLACED. AFTER REPAIR, THE DEVICE WAS FOUND TO PASS ALL DELIVERY, ACCURACY AND FUNCTIONAL TESTS.

Description of Event or Problem · 1

A REPORT WAS RECEIVED THAT STATED THE PUMP ALARMED "CHECK CLUTCH." NO PATIENT INJURY OR ADVERSE EVENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
24118 MEDFUSION 3500 SYRINGE INFUSION PUMP SYRINGE INFUSION PUMP FRN SMITHS MEDICAL MD 3500 NA

Patients

Seq Age Sex Outcome Treatment
1 NONE