FDA Adverse Event
Malfunction
Summary report: N
MEDFUSION 3500 SYRINGE INFUSION PUMP
MDR report key: 2961752
·
Received January 16, 2013
Report
- Report Number
- 2183502-2012-00672
- Event Type
- Malfunction
- Date Received
- January 16, 2013
- Date of Event
- December 17, 2012
- Report Date
- January 15, 2013
- Manufacturer
- SMITHS MEDICAL MD
- Product Code
- FRN
- PMA / PMN Number
- K040899
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION: THE SUSPECT DEVICE WAS RETURNED AND EVALUATED. TESTING WAS ABLE TO REPRODUCE THE CUSTOMER'S COMPLAINT OF A "CHECK CLUTCH" ALARM. INSPECTION OF THE PUMP FOUND MISALIGNED CLUTCH HALVES. FOLLOWING REPLACEMENT OF THE CLUTCH HALVES, THE DEVICE WAS FOUND TO PASS ALL DELIVERY, ACCURACY AND FUNCTIONAL TESTING.
Description of Event or Problem · 1
A REPORT WAS RECEIVED THAT STATED THE PUMP ALARMED "CHECK CLUTCH". NO PATIENT INJURY OR ADVERSE EVENT WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 25163 | MEDFUSION 3500 SYRINGE INFUSION PUMP | SYRINGE INFUSION PUMP | FRN | SMITHS MEDICAL MD | 3500 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NONE |