FDA Adverse Event Malfunction Summary report: N

MEDFUSION 3500 SYRINGE INFUSION PUMP

MDR report key: 2961752 · Received January 16, 2013

Report

Report Number
2183502-2012-00672
Event Type
Malfunction
Date Received
January 16, 2013
Date of Event
December 17, 2012
Report Date
January 15, 2013
Manufacturer
SMITHS MEDICAL MD
Product Code
FRN
PMA / PMN Number
K040899
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE SUSPECT DEVICE WAS RETURNED AND EVALUATED. TESTING WAS ABLE TO REPRODUCE THE CUSTOMER'S COMPLAINT OF A "CHECK CLUTCH" ALARM. INSPECTION OF THE PUMP FOUND MISALIGNED CLUTCH HALVES. FOLLOWING REPLACEMENT OF THE CLUTCH HALVES, THE DEVICE WAS FOUND TO PASS ALL DELIVERY, ACCURACY AND FUNCTIONAL TESTING.

Description of Event or Problem · 1

A REPORT WAS RECEIVED THAT STATED THE PUMP ALARMED "CHECK CLUTCH". NO PATIENT INJURY OR ADVERSE EVENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
25163 MEDFUSION 3500 SYRINGE INFUSION PUMP SYRINGE INFUSION PUMP FRN SMITHS MEDICAL MD 3500 NA

Patients

Seq Age Sex Outcome Treatment
1 NONE