FDA Adverse Event
Malfunction
Summary report: N
9800
MDR report key: 2961645
·
Received February 11, 2013
Report
- Report Number
- 1720753-2013-01653
- Event Type
- Malfunction
- Date Received
- February 11, 2013
- Date of Event
- January 29, 2013
- Report Date
- February 11, 2013
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE BATTERY PACK WAS REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THE SYSTEM DISPLAYED A PRE-CHARGE ERROR CODE. THIS WILL RESULT IN A NO BOOT SITUATION AND THEREBY RESULT IN LOSS OF IMAGING FUNCTIONALITY. THERE ARE NO REPORTS OF PT INJURY OR DEATH ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 58931 | 9800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9800 |