MDR report key: 2961645 · Received February 11, 2013

Report

Report Number
1720753-2013-01653
Event Type
Malfunction
Date Received
February 11, 2013
Date of Event
January 29, 2013
Report Date
February 11, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE BATTERY PACK WAS REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE SYSTEM DISPLAYED A PRE-CHARGE ERROR CODE. THIS WILL RESULT IN A NO BOOT SITUATION AND THEREBY RESULT IN LOSS OF IMAGING FUNCTIONALITY. THERE ARE NO REPORTS OF PT INJURY OR DEATH ASSOCIATED WITH THIS EVENT.