FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 2961609 · Received February 11, 2013

Report

Report Number
1720753-2013-01633
Event Type
Malfunction
Date Received
February 11, 2013
Date of Event
January 28, 2013
Report Date
February 11, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REPRESENTATIVE EVALUATED THE SYSTEM AND REPLACED THE GENERATOR DRIVER BOARD AND A RIBBON CABLE. THE SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SYSTEM DISPLAYED AN OVERLOAD FAULT ERROR MESSAGE. THIS ERROR MAY CAUSE THE SYSTEM TO SHUT DOWN. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THE EVENT DESCRIBED IN THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
59619 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1