FDA Adverse Event Malfunction Summary report: N

DUAL COOLER/HEATER

MDR report key: 2961556 · Received February 11, 2013

Report

Report Number
1828100-2013-00099
Event Type
Malfunction
Date Received
February 11, 2013
Date of Event
January 16, 2013
Report Date
January 16, 2013
Manufacturer
TERUMO CARDIOVASCULAR SYSTEMS CORP.
Product Code
DWC
PMA / PMN Number
PREAMEND
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THIS MDR IS RELATED TO MDR #1828100-2013-00098. THE FIELD SVC REP (FSR) FOUND THAT THE SPEED CONTROL RESISTOR R2 CONTACT WIPER WAS WORN AND WILL NOT ADJUST TO 35.

Description of Event or Problem · 1

THE FIELD SVC REP (FSR) REPORTED THAT DURING PREVENTATIVE MAINTENANCE (PM) OF THE DEVICE, THE WATER PUMP DOES NOT RUN IN "COOL #2" MODE. SINCE THE EVENT OCCURRED DURING PREVENTATIVE MAINTENANCE, THERE WAS NO PT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
59918 DUAL COOLER/HEATER COOLING AND HEATING SYSTEM DWC TERUMO CARDIOVASCULAR SYSTEMS CORP. 11160

Patients

Seq Age Sex Outcome Treatment