FDA Adverse Event Malfunction Summary report: N

DUAL COOLER/HEATER

MDR report key: 2961538 · Received February 11, 2013

Report

Report Number
1828100-2013-00097
Event Type
Malfunction
Date Received
February 11, 2013
Date of Event
January 16, 2013
Report Date
January 16, 2013
Manufacturer
TERUMO CARDIOVASCULAR SYSTEMS CORP.
Product Code
DWC
PMA / PMN Number
PREAMEND
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE FIELD SVC REP (FSR) CONFIRMED THE REPORTED ISSUE. THE FSR REPLACED THE 30/40 DEGREE THERMOMETER. HE VERIFIED THE NEW TEMPERATURE CALIBRATION POINTS. WITH THE PART REPLACED, THE UNIT OPERATED TO MFR SPECIFICATIONS AND WAS RETURNED TO CLINICAL USE. IF ADD'L INFO BECOMES AVAILABLE ON THIS COMPLAINT THAT WOULD ALTER THE FACTS AND/OR CONCLUSION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.

Description of Event or Problem · 1

THE FIELD SVC REP (FSR) REPORTED THAT DURING PREVENTATIVE MAINTENANCE (PM) OF THE DEVICE, THE COOLER HEATER UNIT WAS FAILING 40 DEGREE CELSIUS AND 42 DEGREE CELSIUS THERMOSTAT SHUTOFF TESTS. THE ACTUAL MEASURED TEMPERATURE WERE 42.5 DEGREE CELSIUS AND 44 DEGREE CELSIUS RESPECTIVELY. SINCE THE EVENT OCCURRED DURING PREVENTATIVE MAINTENANCE, THERE WAS NO PT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
59870 DUAL COOLER/HEATER COOLING AND HEATING SYSTEM DWC TERUMO CARDIOVASCULAR SYSTEMS CORP. 11160

Patients

Seq Age Sex Outcome Treatment
1