GYNECARE MORCELLEX* TISSUE MORCELLATOR
Report
- Report Number
- 2210968-2013-01015
- Event Type
- Malfunction
- Date Received
- February 14, 2013
- Date of Event
- January 10, 2013
- Report Date
- January 21, 2013
- Manufacturer
- ETHICON, INC.
- Product Code
- HET
- PMA / PMN Number
- K100280
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). CONCLUSION: THE PRODUCT UPON WHICH THIS MEDWATCH IS BASED IS ANTICIPATED. ONCE THE PRODUCT IS RECEIVED, ANY FURTHER INFORMATION DERIVED FROM THE EVALUATION WILL BE SUBMITTED IN A SUPPLEMENTAL 3500A FORM.
(B)(4): THE ACTUAL DEVICE INVOLVED IN THIS EVENT WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY AND FUNCTIONALLY EVALUATED. THE BLADE FAILED TO ROTATE DURING THE EVALUATION.
IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC ASSISTED VAGINAL HYSTERECTOMY ON (B)(6) 2013. DURING THE PROCEDURE, THE SURGEON GRASPED THE TRIGGER AND THE BLADE DID NOT ROTATE THOUGH IT CAME OUT FROM THE SHEATH. IT WAS UNKNOWN HOW THE PROCEDURE WAS COMPLETED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES. THE SURGEON OPINED THAT THE TISSUE MAY HAVE BEEN CAUGHT BETWEEN THE BLADE AND THE SHEATH.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 66089 | GYNECARE MORCELLEX* TISSUE MORCELLATOR | LAPAROSCOPE, GYNECOLOGIC | HET | ETHICON, INC. | NA | NI |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |