FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 3000 SYSTEM
MDR report key: 2961458
·
Received January 18, 2013
Report
- Report Number
- 3008642652-2013-00197
- Event Type
- Malfunction
- Date Received
- January 18, 2013
- Date of Event
- December 18, 2012
- Report Date
- January 17, 2013
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVAL SUMMARY: DEVICE EVAL OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (BENT PINS OR DAMAGED ELECTRODE BELT CONNECTOR) WAS CONFIRMED. AS REC'D, THE TRUNK CABLE CONNECTOR HOUSING WAS BROKEN. THE CONNECTOR LOCKING NUT WAS MISSING, WHICH WOULD PREVENT THE ELECTRODE BELT FROM PROPERLY SECURING INTO THE MONITOR. THE ROOT CAUSE FOR THE MISSING LOCKING NUT CANNOT BE POSITIVELY IDENTIFIED BUT IS LIKELY PHYSICAL ABUSE. NO ADVERSE EVENT RESULTED FROM THE DAMAGED ELECTRODE BELT CONNECTOR. THE PT REC'D A REPLACEMENT ELECTRODE BELT.
Description of Event or Problem · 1
A (B)(6) MALE PT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT HIS ELECTRODE BELT CONNECTOR WAS DAMAGED. THE PT REC'D A REPLACEMENT ELECTRODE BELT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 28312 | LIFEVEST WCD 3000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 3000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 72 YR |