LIFEVEST WCD 4000 SYSTEM
Report
- Report Number
- 3008642652-2013-00186
- Event Type
- Malfunction
- Date Received
- January 18, 2013
- Date of Event
- December 31, 2012
- Report Date
- January 16, 2013
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- NURSE
Narratives
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF MONITOR SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (CODE 102) WAS CONFIRMED. THE CAUSE FOR THE CODE 102 IS AN OPEN RESISTOR R45 ON THE COMPUTER/ANALOG (CA) BOARD. THE CAUSE FOR THE OPEN RESISTOR IS EXCESSIVE CURRENT. THE CAUSE FOR THE EXCESSIVE CURRENT IS A BROKEN NEGATIVE LEA ON HIGH-VOLTAGE CAPACITOR C21 ON THE DEFIBRILLATOR BOARD. THE EXCESSIVE CURRENT ALSO SHORTED COMPONENTS U15-U18 ON THE DEFIBRILLATOR BOARD. THE ROOT CAUSE FOR THE BROKEN LEAD ON C21 COULD NOT BE POSITIVELY IDENTIFIED, BUT THE DAMAGE LIKELY RESULTED FROM THE MONITOR IMPACTING A HARD SURFACE. ROOT CAUSE ANALYSIS OF SIMILAR EVENTS FOUND THAT EXCESSIVE STRAIN ON THE PRINTED CIRCUIT ASSEMBLY (PCA) FROM SEVERE IMPACT CAN CAUSE FRACTURED SOLDER CONNECTIONS. (B)(4). NO ADVERSE EVENT RESULTED FROM THE BROKEN LEAD ON C21. THE PATIENT RECEIVED A REPLACEMENT MONITOR.
AN (B)(6) FEMALE PATIENT'S NURSE CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT A CODE 102 - CHARGE PROFILE FAULT. THE PATIENT WAS ISSUED A REPLACEMENT MONITOR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 27751 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 80 YR |