FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 2961395 · Received January 18, 2013

Report

Report Number
3008642652-2013-00186
Event Type
Malfunction
Date Received
January 18, 2013
Date of Event
December 31, 2012
Report Date
January 16, 2013
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF MONITOR SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (CODE 102) WAS CONFIRMED. THE CAUSE FOR THE CODE 102 IS AN OPEN RESISTOR R45 ON THE COMPUTER/ANALOG (CA) BOARD. THE CAUSE FOR THE OPEN RESISTOR IS EXCESSIVE CURRENT. THE CAUSE FOR THE EXCESSIVE CURRENT IS A BROKEN NEGATIVE LEA ON HIGH-VOLTAGE CAPACITOR C21 ON THE DEFIBRILLATOR BOARD. THE EXCESSIVE CURRENT ALSO SHORTED COMPONENTS U15-U18 ON THE DEFIBRILLATOR BOARD. THE ROOT CAUSE FOR THE BROKEN LEAD ON C21 COULD NOT BE POSITIVELY IDENTIFIED, BUT THE DAMAGE LIKELY RESULTED FROM THE MONITOR IMPACTING A HARD SURFACE. ROOT CAUSE ANALYSIS OF SIMILAR EVENTS FOUND THAT EXCESSIVE STRAIN ON THE PRINTED CIRCUIT ASSEMBLY (PCA) FROM SEVERE IMPACT CAN CAUSE FRACTURED SOLDER CONNECTIONS. (B)(4). NO ADVERSE EVENT RESULTED FROM THE BROKEN LEAD ON C21. THE PATIENT RECEIVED A REPLACEMENT MONITOR.

Description of Event or Problem · 1

AN (B)(6) FEMALE PATIENT'S NURSE CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT A CODE 102 - CHARGE PROFILE FAULT. THE PATIENT WAS ISSUED A REPLACEMENT MONITOR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
27751 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 80 YR