FDA Adverse Event
Malfunction
Summary report: N
840 VENTILATOR
MDR report key: 2961350
·
Received February 11, 2013
Report
- Report Number
- 8020893-2013-00331
- Event Type
- Malfunction
- Date Received
- February 11, 2013
- Date of Event
- January 14, 2013
- Report Date
- January 14, 2013
- Manufacturer
- COVIDIEN, FORMERLY NELLCOR
- Product Code
- CBK
- PMA / PMN Number
- K970460
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- NOT APPLICABLE
Narratives
Description of Event or Problem · 1
THE SERVICE REPORT SHOWS THE CUSTOMER REPORTED THAT THE 840 VENTILATOR STOPPED CYCLING WHILE IN USE ON A PT. THE PT WAS NOT HARMED OR INJURED AS A RESULT OF THE EVENT. THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION. THE CSE INSPECTED THE DEVICE AND REPLACED THE EXHALATION MODULE PCB AND O2 SENSOR. THE UNIT PASSED EXTENDED SELF TESTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 60004 | 840 VENTILATOR | VENTILATOR | CBK | COVIDIEN, FORMERLY NELLCOR | 840 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |