FDA Adverse Event Malfunction Summary report: N

COYOTE¿ ES

MDR report key: 2961180 · Received February 14, 2013

Report

Report Number
2134265-2013-00691
Event Type
Malfunction
Date Received
February 14, 2013
Date of Event
January 21, 2013
Report Date
January 21, 2013
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
LIT
PMA / PMN Number
K080982
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MANUFACTURER: EXAMINATION OF THE RETURNED DEVICE REVEALED BLOOD IN THE BALLOON AND INFLATION LUMEN. MAGNIFIED INSPECTION REVEALED NO DAMAGE. WHEN POSITIVE PRESSURE WAS APPLIED WITH AN INFLATION DEVICE FILLED WITH WATER, A STREAM OF WATER EMITTED FROM A PINHOLE ALIGNED WITH THE MARKERBAND. THE BALLOON PRESENTED NO IRREGULARITIES IN THE BALLOON MATERIAL AND THE RO MARKER THAT COULD HAVE CONTRIBUTED TO THE DAMAGE. THERE WAS NO EVIDENCE OF ANY PRODUCT QUALITY DEFICIENCIES CONTRIBUTING TO THE DAMAGE. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL/PROCEDURAL FACTORS. (B)(4).

Additional Manufacturer Narrative · 1

AGE AT TIME OF EVENT: 18 YEARS OR OLDER. (B)(4).

Description of Event or Problem · 1

IT WAS FURTHER REPORTED THAT THE TARGET LESION WAS DENOVO LOCATED IN THE SEVERELY CALCIFIED VESSEL BELOW THE KNEE. THE COYOTE ES BALLOON CATHETER WAS REMOVED INTACT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY TREATMENT PROCEDURE A BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE CALCIFIED AND MODERATELY TORTUOUS VESSEL LOCATED BELOW THE KNEE. A 1.5MM X 20MM X 143CM COYOTE ES MR BALLOON CATHETER WAS ADVANCED TO THE TARGET LESION. DURING THE FIRST INFLATION AT 4ATMS A BALLOON RUPTURE OCCURRED. THE COYOTE ES WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS IS LISTED AS GOOD

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
65906 COYOTE¿ ES CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL LIT BOSTON SCIENTIFIC - MAPLE GROVE H74939135152010 15397422

Patients

Seq Age Sex Outcome Treatment
1