FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 2961171 · Received February 14, 2013

Report

Report Number
1416980-2013-03743
Event Type
Malfunction
Date Received
February 14, 2013
Date of Event
February 1, 2013
Report Date
February 1, 2013
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE PROBLEM WAS CONFIRMED. THE ROOT CAUSE WAS LOOSE CONNECTION. THIS ISSUE IS BEING INVESTIGATED THROUGH CAPA. PER THE CUSTOMER THE SAMPLE WAS DISCARDED AND THE LOT NUMBER WAS UNKNOWN, THEREFORE NO EVALUATION OR BATCH REVIEW COULD BE PERFORMED. A FOLLOW-UP REPORT WILL BE FILED IF ANY ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

THE CUSTOMER CONTACTED BAXTER'S TECHNICAL SERVICE CENTER REGARDING A SYSTEM ERROR (SE) 2240 ALARM, WHICH OCCURRED ON THE HOMECHOICE (HC) DURING DWELL 3 OF 4. THE HOME PATIENT (HP) DID NOT DISCONNECT AND NEITHER DID A BAG. THERE WAS SOLUTION IN THE HEATER BAG ONLY AND THE LUER ON THE SUPPLY LINE WAS NOT COMPLETELY TIGHT. THEY CYCLED POWER TO SE 2367 AND THE HP HAS NO MANUAL SUPPLIES AND WOULD CALL THE REGISTERED NURSE (RN). THE ALARM WAS CLEARED. THE PATIENT WAS CONNECTED EITHER AT THE TIME OF THE ALARM OR OBSERVED AIR. THE PATIENT LINE WAS PROPERLY PRIMED BEFORE CONNECTING. PATIENT EXTENSION LINES WERE NOT USED. THE PATIENT DID NOT DISCONNECT PRIOR TO THE ALARM OR OBSERVED AIR. THE SUPPLIES WERE NOT DAMAGED BY AN OUTLET PORT CLAMP OR AN ASSIST DEVICE USED TO MAKE THE CONNECTIONS. PROPER PROCEDURES PER THE USER MANUAL WERE REVIEWED WITH THE REPORTER. IT WAS UNKNOWN HOW THE HP WOULD COMPLETE THERAPY. THERE WAS PATIENT INVOLVEMENT BUT NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE INITIAL REPORT. NO ADDITIONAL INFORMATION WAS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
65495 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 59 YR HOME CHOICE