FDA Adverse Event Malfunction Summary report: N

CS300

MDR report key: 2961163 · Received January 30, 2013

Report

Report Number
2249723-2013-00011
Event Type
Malfunction
Date Received
January 30, 2013
Date of Event
December 11, 2012
Report Date
December 11, 2012
Manufacturer
DATASCOPE CORP
Product Code
DSP
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THE COMPANY REPRESENTATIVE REPLACED THE SOLENOID DRIVER EXCHANGE PCB (PART NUMBER: 0670-00-0639). THE IABP WAS TESTED TO FACTORY SPECIFICATION. IT FUNCTIONED NORMALLY AND WAS RETURNED TO THE CUSTOMER. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT WHILE THE IABP WAS IN USE ON A PT, THE IABP SHUTDOWN. THE PT WAS SWITCHED TO ANOTHER IABP AND THERAPY WAS CONTINUED. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
41300 CS300 INTRA-AORTIC BALLOON PUMP DSP DATASCOPE CORP CS300 NA

Patients

Seq Age Sex Outcome Treatment
1 NI