FDA Adverse Event Malfunction Summary report: N

PUMP IN STYLE ADVANCED BREAST PUMP METRO BAG

MDR report key: 2961154 · Received January 30, 2013

Report

Report Number
1419937-2013-00055
Event Type
Malfunction
Date Received
January 30, 2013
Date of Event
January 2, 2013
Report Date
January 2, 2013
Manufacturer
MEDELA, INC.
Product Code
HGX
PMA / PMN Number
K031614
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NE, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

A REPLACEMENT TRANSFORMER WAS SENT TO THE CUSTOMER. ATTEMPTS TO RETRIEVE THE PRODUCT WERE UNSUCCESSFUL. THE PRODUCT INVOLVED IN THE COMPLAINT WAS NOT RETURNED FOR TESTING/ANALYSIS. THEREFORE, NO CONCLUSIONS CAN BE MADE AS TO THE CAUSE OF THE EVENT. HOWEVER, THE CUSTOMER REPORTED THE HOUSING CAME APART AND THE INNER CIRCUITRY WAS EXPOSED, WHICH IS A SAFETY RISK. THIS TYPE OF FAILURE IS TYPICALLY ASSOCIATED WITH DROPPING THE UNIT ONTO A HARD SURFACE OR OTHER TYPE OF SEVERE IMPACT. THE ORIGINAL DESIGN TESTING INVOLVED DROPPING THE UNIT FROM A 1.0 M HEIGHT PER UL2601-1 2ND EDITION. PRODUCTION SAMPLES WERE DROPPED THREE TO FIVE TIMES FROM A HEIGHT OF 1.0 M AND THE HOUSINGS SHOWED DAMAGE AND CRACKING (ONLY AFTER AT LEAST THREE DROPS). THIS PRODUCT WAS RELEASED AS A RESULT OF CAPA (B)(4) WHICH WAS INITIATED FOR PUMP IN STYLE TRANSFORMER OVERHEATING/MELTING ISSUES FOR WHICH A FIELD ACTION SAFETY NOTIFICATION WAS INITIATED ON 04/04/2011. COMPLAINTS AGAINST THIS PRODUCT ARE CURRENTLY BEING INVESTIGATED (B)(4) IN ORDER TO DETERMINE ROOT CAUSE AND WILL CONTINUE TO BE MONITORED. SHOULD ADDITIONAL INFO OR THE ORIGINAL PRODUCT BE RECEIVED, RESULTING IN NEW, CHANGED, OR CORRECTED INFO, A F/U REPORT WILL BE FILED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED TO CUSTOMER SERVICE THAT HOUSING BROKE OFF WHEN SHE UNPLUGGED IT FROM THE WALL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
41298 PUMP IN STYLE ADVANCED BREAST PUMP METRO BAG NONE HGX MEDELA, INC. 57036/9207010 REV L - 2011

Patients

Seq Age Sex Outcome Treatment
1