FDA Adverse Event
Malfunction
Summary report: N
TOTALCARE BED
MDR report key: 2961152
·
Received January 17, 2013
Report
- Report Number
- 1824206-2013-00484
- Event Type
- Malfunction
- Date Received
- January 17, 2013
- Date of Event
- January 3, 2013
- Report Date
- January 3, 2013
- Manufacturer
- HILL-ROM INC
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MS, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE TECHNICIAN HAD REPLACED THE HEAD UP AND DOWN VALVES AND THE ISSUE REMAINED. THE TECHNICIAN REPLACED THE HYDRAULIC MANIFOLD TO RESOLVE THE ISSUE.
Description of Event or Problem · 1
THE TECHNICIAN ALLEGED THAT THE HEAD OF THE BED WAS DRIFTING DOWN SLOWLY. NO PT IMPACT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 25754 | TOTALCARE BED | A/C POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM INC | 1900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |