FDA Adverse Event Malfunction Summary report: N

TOTALCARE BED

MDR report key: 2961152 · Received January 17, 2013

Report

Report Number
1824206-2013-00484
Event Type
Malfunction
Date Received
January 17, 2013
Date of Event
January 3, 2013
Report Date
January 3, 2013
Manufacturer
HILL-ROM INC
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MS, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN HAD REPLACED THE HEAD UP AND DOWN VALVES AND THE ISSUE REMAINED. THE TECHNICIAN REPLACED THE HYDRAULIC MANIFOLD TO RESOLVE THE ISSUE.

Description of Event or Problem · 1

THE TECHNICIAN ALLEGED THAT THE HEAD OF THE BED WAS DRIFTING DOWN SLOWLY. NO PT IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
25754 TOTALCARE BED A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM INC 1900

Patients

Seq Age Sex Outcome Treatment
1