FDA Adverse Event Malfunction Summary report: N

TOTALCARE BARIATRIC

MDR report key: 2961147 · Received January 17, 2013

Report

Report Number
1824206-2013-00482
Event Type
Malfunction
Date Received
January 17, 2013
Date of Event
December 21, 2012
Report Date
December 22, 2012
Manufacturer
HILL-ROM INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NV, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE HILL-ROM TECHNICIAN FOUND THE CABLING TO PAG BOARD WAS REVERSED. HE CONNECTED THE CABLES CORRECTLY TO RESOLVE THE ISSUE.

Description of Event or Problem · 1

INFORMATION RECEIVED INDICATES WHEN THE INTELLIDRIVE IS ACTIVATED IT RUNS IN REVERSE. THE INTELLIDRIVE IS AN ATTACHED POWER DRIVE MECHANISM BUILT INTO THE BED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
26953 TOTALCARE BARIATRIC AC POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM INC. 1840

Patients

Seq Age Sex Outcome Treatment
1