HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Report
- Report Number
- 1416980-2013-03739
- Event Type
- Malfunction
- Date Received
- February 14, 2013
- Date of Event
- January 31, 2013
- Report Date
- February 1, 2013
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K923065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- PATIENT
Narratives
(B)(4). THE COMPLAINT WAS CONFIRMED. THE CAUSE OF THE SYSTEM ERROR 2240 WAS USE ERROR. THERE WAS NO ALLEGATION REPORTED AGAINST THE BAXTER PRODUCT BY THE CUSTOMER; THEREFORE, THE SAMPLE WAS NOT REQUESTED FOR EVALUATION AND A SAMPLE EVALUATION WILL NOT BE CONDUCTED. THE LOT NUMBER WAS NOT PROVIDED, THEREFORE NO BATCH REVIEW COULD BE PERFORMED.
THE CUSTOMER CONTACTED BAXTER'S TECHNICAL SERVICE CENTER REGARDING A SYSTEM ERROR (SE) 2240/2367 ALARM, WHICH OCCURRED ON THE HOMECHOICE (HC) DURING USE. THE HOME PATIENT (HP) STATED THERE WAS AN UNUSED LINE IN THE ORGANIZER WITH THE CLAMP OPEN. THE BAXTER TECHNICAL SERVICE REPRESENTATIVE (TSR) ASSISTED THE HP TO END THERAPY. THE PATIENT WAS CONNECTED EITHER AT THE TIME OF THE ALARM OR OBSERVED AIR. THE PATIENT LINE WAS PROPERLY PRIMED BEFORE CONNECTING. PATIENT EXTENSION LINES WERE NOT USED. THE PATIENT DID NOT DISCONNECT PRIOR TO THE ALARM OR OBSERVED AIR. ALL BAGS WERE PROPERLY CONNECTED. THE SUPPLIES WERE NOT DAMAGED BY AN OUTLET PORT CLAMP OR AN ASSIST DEVICE USED TO MAKE THE CONNECTIONS. PROPER PROCEDURES PER THE USER MANUAL WERE REVIEWED WITH THE REPORTER. IT WAS UNKNOWN HOW THE HP WOULD COMPLETE THERAPY. THERE WAS PATIENT INVOLVEMENT BUT NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE INITIAL REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 65584 | HOMECHOICE AUTOMATED PD SET WITH CASSETTE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 49 YR | HOME CHOICE |