FDA Adverse Event Malfunction Summary report: N

TRILOGY 202

MDR report key: 2960951 · Received January 30, 2013

Report

Report Number
2518422-2013-00128
Event Type
Malfunction
Date Received
January 30, 2013
Date of Event
January 9, 2013
Report Date
January 9, 2013
Manufacturer
RESPIRONICS, INC.
Product Code
CBK
PMA / PMN Number
K093905
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DURING THE EVAL OF THE DEVICE AT THE MFR'S SERVICE CENTER, A "SERVICE REQUIRED" CODE WAS FOUND IN THE VENTILATOR'S DOWNLOADED ERROR LOG. THE DEVICE'S OXYGEN BLENDING MODULE BOARD WAS REPLACED TO ADDRESS THE ISSUE.

Description of Event or Problem · 1

THE MFR RECEIVED INFO ALLEGING A "SERVICE REQUIRED" ALARM CONDITION OCCURRED. THERE WAS NO HARM OR INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
40506 TRILOGY 202 VENTILATOR, CONTINUOUS, FACILITY USE CBK RESPIRONICS, INC. 1040007

Patients

Seq Age Sex Outcome Treatment
1