FDA Adverse Event
Malfunction
Summary report: N
9800
MDR report key: 2960892
·
Received January 30, 2013
Report
- Report Number
- 1720753-2013-01269
- Event Type
- Malfunction
- Date Received
- January 30, 2013
- Date of Event
- January 22, 2013
- Report Date
- January 30, 2013
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE HARD DRIVE WAS REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THE SYSTEM MIXED PATIENT IMAGE DATA. THERE IS NO REPORT OF PT INJURY OR DEATH.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 42009 | 9800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |