FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 2960856 · Received January 30, 2013

Report

Report Number
1720753-2013-01287
Event Type
Malfunction
Date Received
January 30, 2013
Date of Event
January 23, 2013
Report Date
January 30, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. ALL FAST STOP SWITCH CABLES WERE RESEATED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE FIELD ENGINEER REPORTED THAT THE SYSTEM DISPLAYED A FAST STOP ACTIVATED ERROR MESSAGE. THIS ERROR CAUSES THE SYSTEM TO SHUTDOWN. THERE IS NO REPORT OF PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
41426 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS 9800

Patients

Seq Age Sex Outcome Treatment
1