FDA Adverse Event Malfunction Summary report: N

MOTOR EM210 LEGEND STYLUS TOUCH

MDR report key: 2960756 · Received February 14, 2013

Report

Report Number
1045254-2013-00102
Event Type
Malfunction
Date Received
February 14, 2013
Report Date
January 28, 2013
Manufacturer
MEDTRONIC XOMED, INC
Product Code
ERL
PMA / PMN Number
K081475
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE RECEIVED: (B)(4) 2013. DRILL PASSED THE VISUAL INSPECTION, ATTACHMENT INSPECTION, DIRECTION <(>&<)> SPEED OF ROTATION TESTS, AND VOLTAGE CHECKS IN INDEXED POSITIONS. DRILL FAILED THE TEMPERATURE TEST. THE MEASURED TEMPERATURE RISE DURING THE EVALUATION WAS 37.7 DEGREES F; THE CRITERIA PER THE XPI STATES THAT THE TEMPERATURE RISE MUST BE LESS THAN OR EQUAL TO 27 DEGREES F. THE HIGHEST DETECTED TEMPERATURE DURING THE EVALUATION WAS 114.4 DEGREES F. CONCLUSION: THE "MOST LIKELY" CAUSE OF THIS EVENT CANNOT BE DETERMINED.

Additional Manufacturer Narrative · 1

THIS DEVICE IS USED FOR THERAPEUTIC PURPOSES. (B)(4). PRODUCT NOT RETURNED FOR EVALUATION. METHOD - NO TESTING METHODS PERFORMED.

Description of Event or Problem · 1

IT WAS REPORTED THE HANDPIECE WAS OVERHEATING/RUNNING HOT/HEATING UP. THERE WAS NO PATIENT IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
65094 MOTOR EM210 LEGEND STYLUS TOUCH DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE ERL MEDTRONIC XOMED, INC EM210 205856168