FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 2960733 · Received January 25, 2013

Report

Report Number
8020893-2013-00208
Event Type
Malfunction
Date Received
January 25, 2013
Date of Event
January 4, 2013
Report Date
January 4, 2013
Manufacturer
COVIDIEN, FORMERLY NELLCOR
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE 840 VENTILATOR HAD A BLANK SCREEN. COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) INSPECTED THE DEVICE. THE DEVICE WAS UPGRADED TO 10.4 GRAPHICAL USER INTERFACE DISPLAY. THE UNIT PASSED EXTENDED SELF-TESTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
36309 840 VENTILATOR VENTILATOR CBK COVIDIEN, FORMERLY NELLCOR 840

Patients

Seq Age Sex Outcome Treatment
1