FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 2960720 · Received January 25, 2013

Report

Report Number
8020893-2013-00193
Event Type
Malfunction
Date Received
January 25, 2013
Date of Event
January 2, 2013
Report Date
January 10, 2013
Manufacturer
COVIDIEN, FORMERLY NELLCOR
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

RECEIVED INFORMATION STATING THAT DUE TO A VENTILATOR MALFUNCTION PATIENT WAS PLACED ON A SECOND VENTILATOR. THE PATIENT WAS NOT HARMED OR INJURED AS A RESULT OF THE EVENT. THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION. THE CSE INSPECTED THE DEVICE AND REPLACED THE OXYGEN SENSOR. THE UNIT PASSED EXTENDED SELF TESTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
35518 840 VENTILATOR VENTILATOR CBK COVIDIEN, FORMERLY NELLCOR 840

Patients

Seq Age Sex Outcome Treatment
1