FDA Adverse Event Malfunction Summary report: N

TENACULUM FORCEPS INSTRUMENT

MDR report key: 2960715 · Received January 25, 2013

Report

Report Number
2955842-2013-00304
Event Type
Malfunction
Date Received
January 25, 2013
Date of Event
December 12, 2012
Report Date
December 28, 2012
Manufacturer
INTUITIVE SURGICAL, INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. THE ALLEGED COMPLAINT THAT THE INSTRUMENT'S TIP WOULD NOT STRAIGHTEN COULD NOT BE CONFIRMED. THE DRIVE CABLES AT THE INSTRUMENT'S WRIST WERE INTACT AND UNDAMAGED. THE WRIST WAS STRAIGHT WHEN INPUT DISC DOGS WERE ALL ALIGNED HORIZONTALLY. THE GRIPS WERE ABLE TO OPEN/CLOSE FINE WHEN INPUT DISCS WERE MANUALLY ROTATED. AN ADDITIONAL OBSERVATION THAT WAS NOT REPORTED BY THE SITE WAS TUBE DAMAGE. THE DISTAL END OF THE MAIN TUBE HAD VARIOUS SCRATCH MARKS WITH LIGHT MATERIAL REMOVAL. THE SCRATCHES WERE .070" - .300" IN LENGTH AND NOT ALIGNED WITH THE TUBE AXIS. EVIDENCE IS NOT CONCLUSIVE, BUT THE DAMAGE MIGHT HAVE BEEN CAUSED BY MISHANDLING/MISUSE. PER THE CUSTOMER, AT THE TIME OF THE REPORTED EVENT, NOTHING FELL INTO A PT DURING A SURGICAL PROCEDURE. THE INSTRUMENTS AND ACCESSORIES USER MANUAL SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS: HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. DO NOT USE AN INSTRUMENT TO CLEAN DEBRIS FROM ANOTHER INSTRUMENT INTRAOPERATIVELY. THIS MAY RESULT IN DAMAGE TO THE INSTRUMENTS OR OTHER UNINTENDED CONSEQUENCES, SUCH AS DISCONNECTION OF THE INSTRUMENT TIP.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI STANDARD TOTAL BENIGN HYSTERECTOMY PROCEDURE, THE SURGICAL STAFF ALLEGED THAT THE TIP OF A TENACULUM FORCEPS INSTRUMENT WOULD NOT STRAIGHTEN. THERE WERE NO MISSING OR FALLEN PIECES REPORTED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PT HARM, ADVERSE OUTCOME, OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
36305 TENACULUM FORCEPS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT, NAY NAY INTUITIVE SURGICAL, INC. 400207-08 M101110818 358

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SURGICAL SYSTEM| ACCESSORIES| ESU