5MM MONOPOLAR CAUTERY INSTRUMENT
Report
- Report Number
- 2955842-2013-00305
- Event Type
- Malfunction
- Date Received
- January 25, 2013
- Date of Event
- December 10, 2012
- Report Date
- December 28, 2012
- Manufacturer
- INTUITIVE SURGICAL, INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY, US
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. THE ALLEGED COMPLAINT OF INSTRUMENT'S CASING BEING TORN WAS CONFIRMED. THE INSTRUMENT'S MAIN TUBE INSULATION APPEARED TO HAVE SCRATCHES AND GOUGE MARKS WITH MATERIAL REMOVED AT THE DISTAL END. EVIDENCE IS NOT CONCLUSIVE, BUT THE DAMAGE WAS LIKELY CAUSED BY EXCESS FORCE CONTACT ON THE MAIN TUBE SURFACE. THE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. DO NOT USE AN INSTRUMENT TO CLEAN DEBRIS FROM ANOTHER INSTRUMENT INTRAOPERATIVELY. THIS MAY RESULT IN DAMAGE TO THE INSTRUMENTS OR OTHER UNINTENDED CONSEQUENCES, SUCH AS DISCONNECTION OF THE INSTRUMENT TIP. PER THE CUSTOMER, AT THE TIME OF THE REPORTED EVENT, NOTHING FELL INTO A PT DURING A SURGICAL PROCEDURE.
IT WAS REPORTED THAT DURING A DA VINCI SI SURGICAL PROCEDURE, THE SURGICAL STAFF ALLEGED THAT THE "CASING" OF A MONOPOLAR CAUTERY INSTRUMENT WAS TORN. THERE WERE NO MISSING OR FALLEN PIECES REPORTED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PT HARM, ADVERSE OUTCOME, OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 35463 | 5MM MONOPOLAR CAUTERY INSTRUMENT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT, NAY | NAY | INTUITIVE SURGICAL, INC. | 420142-04 | S10120525 613 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | ESU| DA VINCI SI SURGICAL SYSTEM| ACCESSORIES |