FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 2960664 · Received January 25, 2013

Report

Report Number
1720753-2013-01086
Event Type
Malfunction
Date Received
January 25, 2013
Date of Event
January 7, 2013
Report Date
January 25, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE SYSTEM INTERFACE BOARD, GENERATOR INTERFACE BOARD, HARD DRIVE, AND THE SOFTWARE UPGRADE WERE INSTALLED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WERE INTERMITTENT COMMUNICATION FAILED ERRORS. THIS ERROR RESULTS IN A SYSTEM LOCK UP, NO BOOT, OR SHUT DOWN SITUATION DEPENDING ON WHEN THE LOSS OF COMMUNICATION OCCURS. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
35457 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1