FDA Adverse Event Injury Summary report: N

ELITE SCROLL MILLENNIUM SYSTEM

MDR report key: 2960541 · Received February 12, 2013

Report

Report Number
1920664-2013-00017
Event Type
Injury
Date Received
February 12, 2013
Date of Event
January 14, 2013
Report Date
January 14, 2013
Manufacturer
BAUSCH & LOMB, INC.
Product Code
HQC
PMA / PMN Number
K961310
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

NUMEROUS ATTEMPTS WERE MADE TO OBTAIN ADDITIONAL INFORMATION, HOWEVER, NO RESPONSE HAS BEEN RECEIVED. THE SYSTEM WAS TESTED ON-SITE BY A BAUSCH AND LOMB SERVICE TECHNICIAN AND WAS FOUND TO FUNCTION PROPERLY. THE TECHNICIAN REPORTED THAT THE CUSTOMER HAD FAILED TO SECURE THE POWER CABLE WITHIN THE MOUNTED POWER CABLE FIXATION HOLDER AFTER THE INSTALLATION OF THE COMPUTER. THE TECHNICIAN REDIRECTED THE POWER CABLE TO THE CORRECT POSITION.

Description of Event or Problem · 1

REPORT RECEIVED FROM (B)(6) STATES: "THE MACHINE SWITCHED OFF DURING SURGERY. SOME IMPACT ON VISION."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
61300 ELITE SCROLL MILLENNIUM SYSTEM HQC BAUSCH & LOMB, INC. CX3200

Patients

Seq Age Sex Outcome Treatment
1 Other