FDA Adverse Event
Injury
Summary report: N
ELITE SCROLL MILLENNIUM SYSTEM
MDR report key: 2960541
·
Received February 12, 2013
Report
- Report Number
- 1920664-2013-00017
- Event Type
- Injury
- Date Received
- February 12, 2013
- Date of Event
- January 14, 2013
- Report Date
- January 14, 2013
- Manufacturer
- BAUSCH & LOMB, INC.
- Product Code
- HQC
- PMA / PMN Number
- K961310
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
NUMEROUS ATTEMPTS WERE MADE TO OBTAIN ADDITIONAL INFORMATION, HOWEVER, NO RESPONSE HAS BEEN RECEIVED. THE SYSTEM WAS TESTED ON-SITE BY A BAUSCH AND LOMB SERVICE TECHNICIAN AND WAS FOUND TO FUNCTION PROPERLY. THE TECHNICIAN REPORTED THAT THE CUSTOMER HAD FAILED TO SECURE THE POWER CABLE WITHIN THE MOUNTED POWER CABLE FIXATION HOLDER AFTER THE INSTALLATION OF THE COMPUTER. THE TECHNICIAN REDIRECTED THE POWER CABLE TO THE CORRECT POSITION.
Description of Event or Problem · 1
REPORT RECEIVED FROM (B)(6) STATES: "THE MACHINE SWITCHED OFF DURING SURGERY. SOME IMPACT ON VISION."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 61300 | ELITE SCROLL MILLENNIUM SYSTEM | HQC | BAUSCH & LOMB, INC. | CX3200 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |