FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2960515 · Received February 14, 2013

Report

Report Number
2531779-2013-01767
Event Type
Malfunction
Date Received
February 14, 2013
Report Date
January 26, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Additional Manufacturer Narrative · 1

FOLLOW-UP #1 DATE OF SUBMISSION 05/21/2013-DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE BLACK BOX SHOWED THAT THE PUMP WAS RUNNING ON THE WRONG TIME/DATE SETTINGS SINCE (B)(6) 2013. TIME/DATE SETTINGS WERE CORRECTED ON (B)(6) 2013. NO TIME LOSS/GAIN WAS OBSERVED. THE PUMP POWERED UP AND APPROPRIATELY DISPLAYED THE ¿VERIFY¿ SCREEN. THE PUMP WAS EXERCISED FOR 5 DAYS ON 0.5UNIT/HOUR RATE AND NO TIME CHANGES OCCURRED DURING TESTING. EVALUATION REVEALED THE INTERNAL CLOCK BATTERY ON THE PCB BOARD HAD FAILED. THE PUMP WOULD NOT RETAIN THE USER PROGRAMMED DATE AND TIME SETTINGS UPON REMOVAL OF THE PRIMARY AA BATTERY. WHEN A NEW AA BATTERY IS INSERTED THE PUMP DISPLAYS THE DEFAULT DATE AND TIME WHICH MUST BE MANUALLY CONFIRMED (OR RESET) BY THE USER IN ORDER TO PROCEED.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A HISTORY/SETTINGS (TIME LOSS/GAIN) ISSUE. THE CURRENT TIME AND DATE: 5:30PM (B)(6) 2013, THE TIME AND DATE ON THS PUMP IS 10:30PM (B)(6) 2013. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
64433 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 83 YR