FDA Adverse Event Malfunction Summary report: N

INFINITI VISION SYSTEM OZIL

MDR report key: 2960485 · Received January 25, 2013

Report

Report Number
2028159-2013-00003
Event Type
Malfunction
Date Received
January 25, 2013
Date of Event
January 1, 2012
Report Date
December 27, 2012
Manufacturer
ALCON - IRVINE TECHNOLOGY CENTER
Product Code
HQC
PMA / PMN Number
K082845
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE COMPANY REPRESENTATIVE EXAMINED THE SYSTEM AND COULD NOT DUPLICATE THE PROBLEM REPORTED. PREVENTIVE MAINTENANCE WAS PERFORMED. THE SYSTEM WAS THEN TESTED AND MET ALL PRODUCT SPECIFICATIONS. THERE WAS NO SAMPLE RETURNED FOR EVALUATION AND NO ADDITIONAL INFORMATION PROVIDED RELATED TO THIS EVENT. FOR THIS REASON, STEPS COULD NOT BE TAKEN TO REPLICATE OR CONFIRM THE REPORTED EVENT. A REVIEW OF COMPLAINTS FOR THE LAST 24 MONTHS DID NOT INDICATE ANY ADDITIONAL SIMILAR REPORTS FOR THIS SYSTEM. THE ROOT CAUSE CANNOT BE DETERMINED CONCLUSIVELY. (B)(4).

Description of Event or Problem · 1

A CUSTOMER REPORTED THAT DURING A PROCEDURE A SYSTEM MESSAGE DISPLAYED THAT COULD NOT BE CLEARED. AN ALTERNATE SYSTEM WAS BROUGHT IN AND USED, BUT ALSO PRESENTED PROBLEMS. ADDITIONAL INFORMATION WAS REQUESTED BUT NONE HAS BEEN RECEIVED TO DATE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
35426 INFINITI VISION SYSTEM OZIL PHACOFRAGMENTATION SYSTEM HQC ALCON - IRVINE TECHNOLOGY CENTER INFINITI OZIL NA

Patients

Seq Age Sex Outcome Treatment
1 LEGACY SYSTEM