FDA Adverse Event Injury Summary report: N

HYDROVIEW INTRAOCULAR LENS

MDR report key: 2960287 · Received February 8, 2013

Report

Report Number
1119279-2013-00036
Event Type
Injury
Date Received
February 8, 2013
Date of Event
January 11, 2013
Report Date
January 11, 2013
Manufacturer
BAUSCH + LOMB
Product Code
HQL
PMA / PMN Number
P990014
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE LENS WAS REQUESTED BUT NOT RETURNED TO BAUSCH + LOMB. DESPITE ATTEMPTS TO OBTAIN THE LENS LOT NUMBER, IT WAS NOT PROVIDED. THEREFORE, A REVIEW OF THE LOT HISTORY RECORD COULD NOT BE PERFORMED. SINCE THE LENS WAS NOT RETURNED IT IS NOT POSSIBLE TO CONFIRM WHETHER THE OPACIFICATION OBSERVED BY THE PHYSICIAN IS DUE TO CALCIFICATION OR SOME OTHER MECHANISM. THE CAUSE OF THE CALCIFICATION PHENOMENON IS MULTI-FACTORIAL. THREE MAJOR TYPES OF CALCIFICATION HAVE BEEN PREVIOUSLY DESCRIBED: THE PRIMARY FORM REFERS TO CALCIFICATION DUE TO ISSUES INHERENT TO THE IOL. THE SECONDARY FORM REFERS TO DEPOSITION OF CALCIUM ONTO THE SURFACE OF THE IOL MOST LIKELY DUE TO ENVIRONMENTAL CIRCUMSTANCES (EG, CHANGES IN THE AQUEOUS SURROUNDING THE IMPLANTED IOL ASSOCIATED WITH PRE-EXISTING OR CONCURRENT DISEASES). BY DEFINITION, IT IS NOT RELATED TO ANY PROBLEM WITH THE IOL ITSELF. THE THIRD FORM IS A FALSE-POSITIVE OR PSEUDO-CALCIFICATION REFERS TO THOSE CASES IN WHICH OTHER PATHOLOGY IS MISTAKEN FOR CALCIFICATION OR FALSE-POSITIVE STAINING FOR CALCIUM. BASED ON THE CURRENT INFORMATION, THE EXACT ROOT CAUSE OF THE IOL OPACIFICATION CANNOT BE DETERMINED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A HYDROVIEW INTRAOCULAR LENS WAS EXPLANTED DUE TO OPACIFICATION. THE LOT AND SERIAL NUMBER WERE NOT PROVIDED, THEREFORE IT HAS NOT BEEN POSSIBLE TO DETERMINE WHETHER THE LENS MODEL IS HYDROVIEW 1.0. ADDITIONAL INFORMATION WAS REQUESTED BUT HAS NOT BEEN RECEIVED TO DATE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
53560 HYDROVIEW INTRAOCULAR LENS HQL / INTRAOCULAR LENS HQL BAUSCH + LOMB H60M

Patients

Seq Age Sex Outcome Treatment
1 Other