FDA Adverse Event Malfunction Summary report: N

UNICEL® DXI 600 ACCESS® IMMUNOASSAY SYSTEM

MDR report key: 2960132 · Received February 13, 2013

Report

Report Number
2122870-2013-00119
Event Type
Malfunction
Date Received
February 13, 2013
Date of Event
January 17, 2013
Report Date
January 17, 2013
Manufacturer
BECKMAN COULTER
Product Code
CEW
PMA / PMN Number
K023764
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

RESULTS:SPINNER/GRIPPER FINGER.

Description of Event or Problem · 1

THE CUSTOMER REPORTED FAILING PARATHYROID HORMONE (PTH) AND PARATHYROID HORMONE - INTRAOPERATIVE (PTH-IO) QUALITY CONTROL (QC) AND CALIBRATIONS INVOLVING A UNICEL DXI 600 ACCESS IMMUNOASSAY SYSTEM. WHILE TROUBLESHOOTING THE ISSUE WITH CUSTOMER TECHNICAL SUPPORT (CTS), A FAILED WASH PORTION OF SYSTEM CHECK WAS DISCOVERED THAT HAD BEEN GENERATED 2 HOURS PRIOR TO THE CALIBRATION AND QC ISSUES. THE CUSTOMER WAS UNAWARE OF THE FAILURE AND HAD CONTINUED TO OPERATE THE INSTRUMENT AND REPORT PATIENT RESULTS OUT OF THE LABORATORY. THE CUSTOMER WAS NO INFORMED OF ANY CHANGE OR DELAY IN PATIENT TREATMENT ASSOCIATED WITH THIS EVENT. THE CTS ADVISED THE CUSTOMER TO REPEAT THEIR SYSTEM CHECKS AGAIN WHICH FAILED TWICE. THE CUSTOMER INDICATED THAT ALL OTHER QC RESULTS HAD BEEN WITHIN RANGE BUT DATA WAS NOT SUPPLIED FOR REVIEW. BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED AND FOUND THE SAMPLE PUMP BELT STRETCHED AND FRAYED. THE FSE REPLACED THE SAMPLE PUMP BELT AND A SUCCESSFUL PRESSURE SENSOR CALIBRATION WAS PERFORMED. THE ASPIRATE PROBE TUBING WAS ALSO REPLACED AND A SPINNER/GRIPPER FINGER WAS FOUND BROKEN. SERVICE WAS PERFORMED TO ADDRESS THE BROKEN SPINNER AND REPAIRS WERE VERIFIED PER ESTABLISHED PROCEDURES. THE LIKELY CAUSE OF THE EVENT WAS DETERMINED TO BE THE ASPIRATE PROBE TUBING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
62382 UNICEL® DXI 600 ACCESS® IMMUNOASSAY SYSTEM ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE CEW BECKMAN COULTER NA NA

Patients

Seq Age Sex Outcome Treatment
1