FDA Adverse Event Malfunction Summary report: N

SERFAS ENERGY SUPER 90-S

MDR report key: 2957914 · Received February 12, 2013

Report

Report Number
0002936485-2013-00026
Event Type
Malfunction
Date Received
February 12, 2013
Date of Event
January 31, 2013
Report Date
January 31, 2013
Manufacturer
STRYKER ENDOSCOPY-SAN JOSE
Product Code
GEI
PMA / PMN Number
K991960
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCT WAS RETURNED FOR INVESTIGATION. THE REPORTED TIP BREAKING WAS CONFIRMED. THE METAL ELECTRODE AND PART OF THE CERAMIC COMPONENT HAD BROKEN OFF FROM THE PROBE¿S TIP. THE BROKEN PARTS WERE ALSO RETURNED FOR EVALUATION. THERE WERE NO VISIBLE WEAR MARKS OR DAMAGE TO THE PROBE'S INSULATION. HOWEVER, SLIGHT SCRATCHES/MARKS COULD BE OBSERVED ON THE METAL LUMEN. THIS CONDITION IS A KNOWN FAILURE MODE AND THE PROBABLE ROOT CAUSES FOR THE REPORTED CONDITION CAN BE ASSOCIATED, BUT NOT LIMITED TO: NON-CONFORMING COMPONENT, POOR ASSEMBLY PROCESS, AND/OR MISUSE. IN SUM, THE PRODUCT WAS RETURNED FOR INVESTIGATION AND THE REPORTED FAILURE MODE WAS CONFIRMED. THE FAILURE MODE WILL BE MONITORED FOR FUTURE REOCCURRENCE.

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION WILL BE PROVIDED ONCE THE INVESTIGATION HAS BEEN COMPLETED.

Description of Event or Problem · 1

IT WAS REPORTED BY A COMPANY REPRESENTATIVE THAT DURING A PROCEDURE THE TIP OF THE UNIT BROKE OFF INSIDE THE PATIENT. THE TIP OF THE UNIT WAS RETRIEVED AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED.

Description of Event or Problem · 1

IT WAS REPORTED BY A COMPANY REPRESENTATIVE THAT DURING A PROCEDURE THE TIP OF THE UNIT BROKE OFF INSIDE THE PATIENT. THE TIP OF THE UNIT WAS RETRIEVED AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
61685 SERFAS ENERGY SUPER 90-S ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES GEI STRYKER ENDOSCOPY-SAN JOSE UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1