FDA Adverse Event Malfunction Summary report: N

VENTED AUTOFEED HUMIDIFICATION CHAMBER

MDR report key: 2957385 · Received February 12, 2013

Report

Report Number
9611451-2013-00094
Event Type
Malfunction
Date Received
February 12, 2013
Date of Event
January 18, 2013
Report Date
January 24, 2013
Manufacturer
FISHER & PAYKEL HEALTHCARE LIMITED
Product Code
BTT
PMA / PMN Number
K934140
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE COMPLAINT MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBER WAS RETURNED TO FISHER & PAYKEL HEALTHCARE IN (B)(4) AND WAS VISUALLY INSPECTED. RESULTS: VISUAL INSPECTION REVEALED THAT THERE WAS A BREAK IN THE FEEDSET TUBING AT THE CONNECTION TO THE CHAMBER DOME. THE SURFACE OF THE BREAK WAS ROUGH (NOT SMOOTHLY CUT). A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 120914. CONCLUSION: THE DAMAGE APPEARED TO BE CAUSED BY THE TUBE BEING PULLED AWAY FROM THE CHAMBER, POSSIBLY DUE TO THE FEEDSET BEING CAUGHT OR UNDER TENSION. ALL CHAMBERS ARE PRESSURE TESTED BEFORE THEY LEAVE THE PRODUCTION LINE; AND ANY HOLES, LEAKS, OR BREAKS IN THE FEEDSET ARE IDENTIFIED DURING THIS PROCESS. THIS SUGGESTS THAT THE DAMAGE OCCURRED POST PRODUCTION. THE USER INSTRUCTIONS WHICH ACCOMPANY THE MR290 CHAMBER STATE THE FOLLOWING: "SET APPROPRIATE VENTILATOR ALARM." "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE COMPLAINT MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBER WAS RETURNED TO FISHER & PAYKEL HEALTHCARE IN (B)(4) AND WAS VISUALLY INSPECTED. RESULTS: VISUAL INSPECTION REVEALED THAT THERE WAS A BREAK IN THE FEEDSET TUBING AT THE CONNECTION TO THE CHAMBER DOME. THE SURFACE OF THE BREAK WAS ROUGH (NOT SMOOTHLY CUT). A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 120914. CONCLUSION: THE DAMAGE APPEARED TO BE CAUSED BY THE TUBE BEING PULLED AWAY FROM THE CHAMBER, POSSIBLY DUE TO THE FEEDSET BEING CAUGHT OR UNDER TENSION. ALL CHAMBERS ARE PRESSURE TESTED BEFORE THEY LEAVE THE PRODUCTION LINE; AND ANY HOLES, LEAKS, OR BREAKS IN THE FEEDSET ARE IDENTIFIED DURING THIS PROCESS. THIS SUGGESTS THAT THE DAMAGE OCCURRED POST PRODUCTION. THE USER INSTRUCTIONS WHICH ACCOMPANY THE MR290 CHAMBER STATE THE FOLLOWING: - "SET APPROPRIATE VENTILATOR ALARM." - "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." (B)(4).

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE (FPH) FIELD REPRESENTATIVE THAT THE WATER FEEDSET TUBE BECAME DISCONNECTED FROM THE CHAMBER DOME OF AN MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBER AFTER THREE DAYS OF USE. NO PATIENT CONSEQUENCE WAS REPORTED.

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE (FPH) FIELD REPRESENTATIVE THAT THE WATER FEEDSET TUBE BECAME DISCONNECTED FROM THE CHAMBER DOME OF AN MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBER AFTER THREE DAYS OF USE. NO PATIENT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
61848 VENTED AUTOFEED HUMIDIFICATION CHAMBER BTT BTT FISHER & PAYKEL HEALTHCARE LIMITED MR290V 120914

Patients

Seq Age Sex Outcome Treatment
1 RT235 INFANT DUAL-HEATED EVAQUA BREATHING CIRCUIT| RT235 INFANT DUAL-HEATED EVAQUA BREATHING CIRCUIT