FDA Adverse Event Malfunction Summary report: N

FENESTRATED BIPOLAR FORCEPS INSTRUMENT

MDR report key: 2955939 · Received February 11, 2013

Report

Report Number
2955842-2013-00467
Event Type
Malfunction
Date Received
February 11, 2013
Date of Event
January 17, 2013
Report Date
January 17, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT THE PITCH UP CABLE WAS BROKEN AT THE DISTAL CLEVIS HUB. THE CABLE SEGMENT THAT CONTAINS THE CRIMP REMAINED INSTALLED IN THE CLEVIS. THE CLEVIS DOES NOT EXHIBIT ANY DAMAGE OR WEAR MARKS. THE OTHER CABLES AT WRIST WERE NOT DAMAGED. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING REPROCESSING OF THE FENESTRATED BIPOLAR FORCEPS INSTRUMENT, IT WAS NOTED THAT THE WIRE AT THE TIP OF THE INSTRUMENT WAS EXPOSED. THERE WERE NO MISSING OR FALLEN PIECES REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
59772 FENESTRATED BIPOLAR FORCEPS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420205-05 M10120514 744

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SURGICAL SYSTEM, ACCESSORIES AND ESU