FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 2955820
·
Received February 11, 2013
Report
- Report Number
- 3004209178-2013-90975
- Event Type
- Malfunction
- Date Received
- February 11, 2013
- Date of Event
- January 20, 2013
- Report Date
- February 1, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MS
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP ALARMED DURING THE BASIC OCCLUSION TEST AS A RESULT OF A LOOSE DRIVE SUPPORT DISK.
Description of Event or Problem · 1
THE CUSTOMER REPORTED REGARDING ISSUES DURING PRIME/REWIND. THE BLOOD GLUCOSE READING WAS 239MG/DL. THE CALLER STATED THAT INSULIN SQUIRTED OUT DURING THE MANUAL PRIME PROCESS, AND THE INSULIN PUMP ALARMED. ADVISED THE CUSTOMER THAT THE INSULIN PUMP WOULD BE REPLACED. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 58964 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-722NAB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 64 YR |