FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2955816 · Received February 11, 2013

Report

Report Number
3004209178-2013-90961
Event Type
Malfunction
Date Received
February 11, 2013
Date of Event
January 15, 2013
Report Date
January 29, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP ALARMED DURING THE BASIC OCCLUSION TEST AS A RESULT OF A LOOSE DRIVE SUPPORT DISK. THE DEVICE HAD MISSING END CAP STICKER.

Description of Event or Problem · 1

THE CUSTOMER REPORTED REGARDING ISSUES DURING PRIME/REWIND. THE BLOOD GLUCOSE READING WAS 150MG/DL. THE CALLER STATED THAT INSULIN SQUIRTED OUT DURING THE MANUAL PRIME PROCESS, AND THE INSULIN PUMP ALARMED. THE CUSTOMER ALSO STATED THAT THE DEVICE WAS STUCK ON THE PRIME LOOP. ADVISED THE CUSTOMER THAT THE INSULIN PUMP WOULD BE REPLACED. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
59887 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722NAS

Patients

Seq Age Sex Outcome Treatment
1 56 YR