FOXCROSS PTA CATHETER
Report
- Report Number
- 2024168-2013-00789
- Event Type
- Malfunction
- Date Received
- February 11, 2013
- Date of Event
- January 21, 2013
- Report Date
- January 21, 2013
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- LIT
- PMA / PMN Number
- K102705
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. IT IS INDICATED THAT THE DEVICE IS NOT RETURNING FOR EVALUATION; THEREFORE A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. THE RESULTS OF THE QUERY OF SIMILAR INCIDENTS IN THE COMPLAINT HANDLING DATABASE FOR THIS LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE REVIEWED INFORMATION, NO PRODUCT DEFICIENCY WAS IDENTIFIED.
IT WAS REPORTED THAT DURING PRE-DILATATION IN THE MILDLY TORTUOUS/CALCIFIED ILIAC ARTERY FOR TREATMENT OF A 50% STENOSIS, THE FOXCROSS BALLOON WAS INFLATED TO 8 ATMOSPHERES. DURING THE SECOND INFLATION, THE BALLOON RUPTURED AT 8 ATMOSPHERES. PRE-DILATATION WAS COMPLETED USING A NON-ABBOTT BALLOON. THERE WAS NO ADVERSE PATIENT EFFECT AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 58293 | FOXCROSS PTA CATHETER | PERIPHERAL DILATATION CATHETER | LIT | AV-TEMECULA-CT | 777323 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |