FDA Adverse Event Malfunction Summary report: N

ADJ WIRE COLLET 0.7-1.8MM

MDR report key: 2955700 · Received February 11, 2013

Report

Report Number
0001811755-2013-00241
Event Type
Malfunction
Date Received
February 11, 2013
Date of Event
January 30, 2013
Report Date
January 30, 2013
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
KIJ
PMA / PMN Number
K972367
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

IT WAS NOTED DURING FAILURE ANALYSIS THAT WEAR ON THE DRIVESHAFT MAY PRODUCE METAL SHAVINGS. THEREFORE, IT IS LIKELY THE REPORTED EVENT WAS DUE TO DRIVESHAFT SCORING.

Description of Event or Problem · 1

THE ADJ WIRE COLLET 0.7-1.8MM WAS SENT FOR SERVICE AND DURING FAILURE ANALYSIS IT WAS NOTED THAT THERE ARE METAL SHAVINGS BY THE DRIVESHAFT AND LEVER OF THE DEVICE. THERE WAS NO PATIENT INVOLVEMENT AND NO ADVERSE CONSEQUENCES ASSOCIATED WITH THE DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
59392 ADJ WIRE COLLET 0.7-1.8MM INSTRUMENT, SURGICAL, ORTHOPEDIC, DC-POWERED MOTOR AND ACCESSORY/ATTACHMENT KIJ STRYKER INSTRUMENTS-KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1