FDA Adverse Event Injury Summary report: N

CONTOUR CURVED CUTTER STAPLER

MDR report key: 2955652 · Received February 11, 2013

Report

Report Number
3005075853-2013-00539
Event Type
Injury
Date Received
February 11, 2013
Date of Event
January 15, 2013
Report Date
January 16, 2013
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K040038
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS SHOWED THAT THE DEVICE WAS RECEIVED IN GOOD VISUAL CONDITIONS AND WITH A CARTRIDGE RELOAD LOADED IN THE DEVICE. THE RELOAD WAS RECEIVED VOID OF STAPLES, WITH THE WASHER COMPLETELY CUT, WITH THE DRIVERS EXPOSED AND WITH THE KNIFE RECESS BELOW THE CARTRIDGE DECK. THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH AN ENGINEERING SAMPLE RELOAD AND THE DEVICE FIRED, FORMING ALL STAPLES AND CUTTING AS INTENDED. THE CUT LINE WAS COMPLETE, THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. THE DEVICE LOCKOUT AND THE RELOAD LOCKOUT WERE FUNCTIONAL. IT SHOULD BE NOTED THAT ALL DEVICES ARE INSPECTED 100% FOR STAPLE PRESENCE BY AN AUTOMATED VISION SYSTEM, AND ARE VISUALLY INSPECTED 100% AS A FINAL CHECK. IN ADDITION, AT FINISHED GOODS THE DEVICES ARE VISUALLY INSPECTED BASED ON A SAMPLE. THE MANUFACTURING RECORDS WERE REVIEWED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LOW RECTAL CANCER PROCEDURE, THE SURGEON USED THE DEVICE TO STAPLE THE RECTUM AND FOUND THAT THE DISTAL PART WAS NOT CLOSED AFTER THE FIRING. THE SURGEON HAD TO CUT THE ANUS, USED HAND SEWING TO COMPLETE THE PROCEDURE AND MADE THE FISTULA FOR THE PATIENT. THE PATIENT HAS NOT BEEN DISCHARGED YET. THE PATIENT LOST ANUS. ADDITIONAL FOLLOW UP IS BEING CONDUCTED. IF ADDITIONAL DETAILS BECOME AVAILABLE A 3500 A SUPPLEMENTAL WILL BE SENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
58453 CONTOUR CURVED CUTTER STAPLER STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. UNK J4AG9Z

Patients

Seq Age Sex Outcome Treatment
1 66 YR Required Intervention