FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO

MDR report key: 2955533 · Received February 11, 2013

Report

Report Number
2649622-2013-01386
Event Type
Injury
Date Received
February 11, 2013
Report Date
November 15, 2012
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED AND ANALYZED. NO ANOMALIES WERE FOUND. THERE WAS BODY TISSUE/FIBROTIC GROWTH ON THE DISTAL END OF THE ELECTRODE. THE DEFIBRILLATION SUPERIOR VENA CAVA COIL WAS PULLED/STRETCHED/OVERSTRESSED. VISUAL ANALYSIS REVEALED THERE WAS APPARENT EXPLANT DAMAGE.

Description of Event or Problem · 1

IT WAS REPORTED THERE WAS A HIGH IMPEDANCE ALERT ON THE LEAD. THE LEAD WAS REMOVED AND A NEW LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
58858 SPRINT QUATTRO DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 6944-65

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R