CONTROL 3328200 PNEUMATIC FOOT
Report
- Report Number
- 1045254-2013-00086
- Event Type
- Malfunction
- Date Received
- February 11, 2013
- Date of Event
- January 13, 2013
- Report Date
- January 21, 2013
- Manufacturer
- XOMED MFG JACKSONVILLE
- Product Code
- ERL
- PMA / PMN Number
- K791407
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
THIS PRODUCT IS USED FOR THERAPEUTIC PURPOSES. (B)(4). THE DEVICE WAS NOT RETURNED TO THE MANUFACTURER FOR EVALUATION.
DEVICE RETURNED (B)(6) 2013. THE DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION BY THE QUALITY ENGINEERING TEAM. THE PRODUCT ANALYSIS FOUND THERE WAS A PIECE OF TAPE AROUND ONE POINT OF THE CABLE. PRESSURE METER ASSET 1270-J WAS USED TO CHECK THE FOOTSWITCH. THE PEDAL WAS DEPRESSED SEVERAL TIMES; READINGS WERE BETWEEN 9 AND 10.5 PSI. PER XPI-NS 3328200 REV J SECTION 9.7.1.1, READING SHOULD BE BETWEEN 5 AND 11 PSI. WHEN THE PEDAL WAS RELEASED, THE PRESSURE WENT BACK TO ZERO. NO FUNCTIONAL ISSUES WERE IDENTIFIED DURING THE EVALUATION. THE REPORTED COMPLAINT COULD NOT BE CONFIRMED. (B)(4).
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT THE PNEUMATIC FOOT CONTROL UNINTENTIONALLY ACTIVATED. THE EVENT DID NOT OCCUR DURING A PROCEDURE AND THERE WAS NO PATIENT IMPACT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 59612 | CONTROL 3328200 PNEUMATIC FOOT | DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE | ERL | XOMED MFG JACKSONVILLE | 3328200 | 0206139618 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |