FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO

MDR report key: 2955431 · Received February 11, 2013

Report

Report Number
2649622-2013-01464
Event Type
Injury
Date Received
February 11, 2013
Date of Event
October 16, 2012
Report Date
March 12, 2013
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED AND ANALYSIS FOUND NO ANOMALIES. HOWEVER, THE LEAD HELIX WAS COMPRESSED AND BENT. VISUAL ANALYSIS NOTED APPARENT EXPLANT DAMAGE.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD SHOWED EPISODES OF NOISE AND OVERSENSING. THERE WERE ALSO INSTANCES OF HIGH PACE/SENSE IMPEDANCE MEASUREMENTS. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
58912 SPRINT QUATTRO DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 694758

Patients

Seq Age Sex Outcome Treatment
1 00056 YR Hospitalization| R CD1257-40Q COMPETITOR IMPLANTABLE CARDIOVERTER DEF