Report Number
9614453-2013-00294
Event Type
Injury
Date Received
February 11, 2013
Date of Event
December 14, 2012
Report Date
April 16, 2013
Manufacturer
IPG MFG SWITZERLAND
Product Code
LWS
PMA / PMN Number
P980016
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY #(B)(4) PERFORMANCE DATA WAS COLLECTED AND ANALYZED. THE SAVE TO DISK PRIMARY FINDING NOTED DEVICE ALERT LOW BATTERY VOLTAGE RECOMMENDED REPLACEMENT TIME (RRT). TIME OF RRT IN SAVE TO DISK ON (B)(6) 2012 DEVICE RRT<(><<)>=2.62 VOLTS. WEEKLY BATTERY VOLTAGE TREND DATA SHOWS MIN BAT=2.64 TO 2.62 VOLTS BETWEEN (B)(6) 2012. TWO LOW BATTERY VOLTAGE ALERTS ON (B)(6) 2012.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THIS MODEL NUMBER IS NOT APPROVED FOR DISTRIBUTION IN THE U.S., HOWEVER, IT IS SIMILAR TO A DEVICE MARKETED IN THE U.S. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY #(B)(4). THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS OF THE DEVICE FOUND A HIGH CURRENT DRAIN CONDITION. ELECTRICAL ANALYSIS REVEALED THE CAUSE OF THE HIGH CURRENT DRAIN WAS CURRENT LEAKAGE IN THE BATTERY FILTER CAPACITORS.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DEVICE REACHED ELECTIVE REPLACEMENT INDICATOR WITH SHORTER THAN EXPECTED LIFETIME LONGEVITY. THE DEVICE WAS EXPLANTED AND WAS REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
58843 VIRTUOSO VR DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS IPG MFG SWITZERLAND D164VWC

Patients

Seq Age Sex Outcome Treatment
1 00072 YR Hospitalization| R